Keep dry
The same packing stacked up to 5-layers
5
Atmospheric pressure limitation.
106kPa
50kPa
30°C
Temperature limitation
2°C
Humidity limitation.
95%
%
0%
European Representative.
This item is compliant with Medical Device Directive 93/42/EEC of June 14, 1993,
a directive of the European Economic Community.
Chapter 10
CAre / CleAnInG
10.1 System Check
Before using the monitor, do the following:
• check if there is any mechanical damage;
• check all the outer cables, inserted modules and accessories;
• check all the functions of the monitor to make sure that the monitor is in good condition.
If you find any damage on the monitor, stop using the monitor on patient, and contact the biomedical
engineer of the hospital or our Customer Service immediately.
The overall check of the monitor, including the safety check, should be performed only by qualified
personnel once every 6 to 12 month, and each time after fix up.
You should check the synchronism of the defibrillator in the frequency described in the hospital regulations.
At least every 3 months, it should be checked by a qualified customer service technician.
All the checks that need to open the monitor should be performed by qualified customer service technician.
The safety and maintenance check can be conducted by persons from our company. You can obtain the
material about the customer service contract from the local office.
The circuits diagrams, parts lists and calibration instructions of the monitor can be provided by the
manufacturer.
Note
To ensure maximum battery life, it is recommended that, at least once a month, the monitor be run on
battery until it turns itself off and then recharged.
Warning
• If the hospital or agency that is responding to using the monitor does not follow a satisfactory maintenance
schedule, the monitor may become invalid, and the human health may be endangered.
• Refer the battery replacement only to our service technician.
10.2 General Cleaning
Warning
• Before cleaning the monitor or the sensor, make sure that the equipment is switched off and disconnected
from the power line.
48