101
Response time to heart rate meter to change in heart rate
80 to 120bpm
80 to 40bpm
Time to ALARM for tachycardia
Tachycardia ventricular:amplitude =1mV(p-v),heart rate = 206bpm
Ampitude
×1
×0.5
×2
Tachycardia ventricular :amplitude =2mV(p-v),heart rate = 195bpm
Ampitude
×1
×0.5
×2
Pace Pulse
Pulse indicator
Pulse Rejection
2.8 reSPIrAtIon
Method
Differential Input Impedance
Measuring Impedance Range:
Base line Impedance Range:
Bandwidth
Scan speed
Resp Rate
Measuring and Alarm Range
Adult
Neonate/Pediatric
Resolution
Accuracy
Apnea Alarm
Alarm limit range (rpm)
Alarm high limit
Alarm low limit
Alarm limit range Step (rpm)
2.9 nIBP
Method
Mode
Measuring Interval in AUTO Mode
Measuring Period in STAT Mode
Alarm Type
Measuring and alarm Range
Adult Mode
SYS 40 ~ 270 mmHg
DIA
MEAN
Pediatric Mode
SYS
DIA
MEAN
< 8s
< 8s
Average(s)
2.7
3.8
3.8
Average(s)
2.3
3.3
2.2
Pulse is marked if the requirements of ANSI/AAMI EC13:2002,Sect.4.1.4.1 are met:
Amplitude: ±2mV ~ ±700mV
Width:
0.1ms ~ 2ms
Pulse is rejected if the requirements of ANSI/AAMI EC13-2002:Sect.4.1.4.1 are met:
Amplitude: ±2mV ~ ±700mV
Width:
0.1ms ~ 2ms
Impedance between R-F(RA-LL)
>2.5 MΩ
0.3~5.0Ω
0.1 KΩ– 2.5 KΩ
0.3 ~ 2.5 Hz
6.25mm/s, 12.5mm/s, 25mm/s
0 ~ 120 rpm
0 ~ 150 rpm
1 rpm
±2 rpm
10 ~ 40 s
(low limit+1)~150
Adult: 0~(high limit-1) - Pediatric and neonate: 0~(high limit-1)
1
Oscillometric
Manual, Auto, STAT
1, 2, 3, 4, 5, 10, 15, 30, 60, 90,120,240,480,960 Min
5 Min
SYS, DIA, MEAN
10 ~ 215 mmHg
20 ~ 235 mmHg
40 ~ 200 mmHg
10 ~ 150 mmHg
20 ~ 165 mmHg
Range (s)
2.2~ 2.9
3.5~4.0
3.5~4.0
Range (s)
2.0~3.0
3.0~4.0
2.0~3.0