DISPENSE THE FIRST SAMPLE
1.
Follow the screen instructions, then press ACCEPT to
store the first sample weight (liquid).
2. Continue to follow the screen instructions by
dispensing samples and pressing ACCEPT to store
each sample weight (liquid).
3. Once all the sample weights have been dispensed,
the results screen is automatically displayed.
NOTE: The default number of samples is 10, but
can be changed in Setup, from 2 to 30 samples.
4. Press GRAPH to see results plotted. Press RESULT to
return to the result screen. Press CLOSE to return to
the pipette adjustment home screen.
5. To view either the results or graph screens again
from the home screen, press the VIEW RESULT button.
To start a new pipette adjustment process, press
BEGIN PIPETTE ADJUSTMENT.
PAGE 28 OF 96
OPERATION CONTINUED
The statistical quality control (SQC) feature is useful during
various types of process filling operations, to monitor and/
or control the process to eliminate under- and over-filling.
As samples are weighed and stored in the balance, a
trend analysis is developed and can be displayed.
Up to 10 batches with up to 80 samples in a batch
are visible for quality control purposes. Each batch of
samples is shown, with the maximum, minimum, standard
deviation and mean values for each batch. An ongoing
examination of the relative deviation of the samples
along with other statistical data can be viewed.
The filling process operation can be effectively monitored
by observing the results on the view trace screen.
SQC collection can be done manually (manual mode,
requires key press after each step) or automatically
(auto mode, the weights are automatically recorded
when stable).
In the upper left portion of the home screen, select SQC.
EDIT SETTINGS
1. Confirm the default values (Nominal, Inaccuracy,
2. To view/edit the default values, touch the EDIT
3. The Settings screen appears.
• Settings Available: Nominal Target weight,
• Functions Available: RETURN TO APPLICATION
4. To adjust the Nominal Target weight value, touch the
5. Key in the desired Nominal Weight, then press √. The
STATISTICAL QUALITY CONTROL (SQC)
Imprecision, Density, Atmospheric Pressure, Water
Temperature, etc.) are correct.
SETTINGS button.
+Tolerance 1, -Tolerance 1, +Tolerance 2,
-Tolerance 2
NOMINAL button. A numeric input window appears.
display returns to the previous screen with the new
value highlighted.
0721 IH-4737