THE EDIT SETTINGS SCREEN
• Settings available: Water temperature
• Functions available: Return to application
1. To adjust the water temperature value, touch the
WATER TEMP button. The balance calculates water
density based on the water temperature value
entered (internal lookup table). Measure the actual
water temperature using a precision thermometer.
2. A numeric input window appears. Enter the actual
water temperature, then press DONE.
3. The display returns to the previous screen with the
new value highlighted. To return to the density
determination home screen, touch RETURN TO
APPLICATION.
4. To begin the density determination process, press
BEGIN DENSITY DETERMINATION.
DENSITY DETERMINATION STEPS
1.
Weigh the sample in air. Press BEGIN DENSITY
CALCULATION. Follow the screen instructions, then
press ACCEPT to store the dry sample weight ("in air").
2. Weigh the sample submerged in the liquid. Follow
the screen instructions, then press ACCEPT to store
the sample weight (submerged in liquid).
PAGE 22 OF 96
OPERATION CONTINUED
3. Once the necessary weights have been determined,
4. Press BEGIN DENSITY DETERMINATION to reset all the
MEASURING THE DENSITY OF A BUOYANT MATERIAL USING
WATER
1.
2. Confirm the following setups are selected:
• Density type: Solid
• Liquid type: Water
• Porous material: OFF
3. Press DONE to return to the density determination
4. Follow the same procedure as solid material, except
MEASURING THE DENSITY OF A SOLID USING AN AUXILIARY
LIQUID
1. To enable this feature, enter the density setup menu
• Density type: Solid
• Liquid type: Other
• Porous material: OFF.
2. Confirm the default values displayed (liquid density,
3. To edit the default values, touch EDIT SETTINGS and
• Settings available: Liquid density
• Functions available: Return to application
4. To adjust the liquid density value, touch the LIQUID
5. Measure the actual liquid density of the density
6. A numeric input window appears. Key in the density
7. The display returns to the previous screen with the
the density of the sample is displayed in g/cm
(along with the weight in air, weight in water) on the
application screen.
weight values and restart the process.
Press the SETUP icon
home screen.
in step 2, push the sample down into the liquid until
it is fully submerged.
and select the following:
etc.) are correct.
the application settings screen appears.
DENSITY button.
setup and enter it.
in g/cm
, then press √.
3
new value highlighted. To return to the density
determination home screen, touch RETURN TO
APPLICATION.
3
to open settings screen.
0721 IH-4737