DIFFERENTIAL WEIGHING WITH AUTO-PROCESS ON
Differential weighing with auto process ON requires the
initial and final weights to be sequentially gathered for
a single item.
1. To enable this feature, touch the SETUP icon
from the home screen to enter application
setup.
2. Press AUTO PROCESS to turn it ON.
3. Follow the screen instructions to begin.
4. Press ITEM SETTINGS to set up items in memory if
necessary. Use the buttons as necessary.
5. Follow the same steps as shown on page 20. Press
RETURN TO APPLICATION when ready.
6. A message prompts for Item 1 to be placed on
the pan again, then to press ACCEPT to establish
final weight. After ACCEPT is pressed, the difference
between initial and final weights appears in the
Reference Fields.
7. A message prompts to start the process again, if
desired.
PAGE 21 OF 96
OPERATION CONTINUED
The balance can be used to determine an object's density.
Four types of density determination can be made:
• Solids more dense than water
• Solids less dense than water
• Liquid density
• Porous material (impregnated with oil)
A density determination kit is designed to be used
with the balance. Illustrations in this procedure refer
to the density kit. However, you may use whatever
lab apparatus that suits the requirements for density
measurements. The balance software includes a built-in
reference density table for water at temperatures
between 10 ° C and 30 ° C. Review this entire section
before attempting density measurements.
When making density measurements the material
should weigh at least 10 mg on an analytical balance
and 100 mg on a precision balance.
Density can be determined manually (with a key
press after each step) or automatically (weights are
automatically recorded). Before making density
measurements, establish the Application Settings.

MEASURING THE DENSITY OF A SOLID USING WATER (DEFAULT)

1.
2. Confirm the following setups are selected:
• Density type: Solid
• Liquid type: Water
• Porous material: Off
3. Press DONE to return to the density determination

DENSITY DETERMINATION

Press the SETUP icon
determination settings screen.
home screen.
to open the density
0721 IH-4737