FORMULATION RESULTS
1.
Press PRINT to send the results to the enabled COM
part.
2. To close the results window and return to the
Application home screen, press CLOSE.
DIFFERENTIAL WEIGHING
Differential weighing stores weight values so samples
can be dried or processed and the difference in weight
calculated. Up to 99 samples can be stored.
Differential weighing has two operating modes:
• Auto-Process ON, where the initial and final weights
are sequentially gathered for a set of items, with no
need to return to item settings.
• Auto-Process OFF, where initial and final weights are
gathered separately for each item, after each item
is selected in item settings.
Touch the SETUP icon
from the home screen.

ITEM SETTINGS

1.
Press ITEM SETTINGS to set up items in memory.
2. To set up the item, press ADD NEW.
PAGE 20 OF 96
OPERATION CONTINUED
to enter application setup
3. New items are added with default name and weight
information (no initial or final weights).
4. To change the default item name, press EDIT NAME.
The alpha-numeric entry screen appears.
5. Enter the new name, then press √. The display
returns to the previous screen with the new value
highlighted momentarily.
6. Continue adding items as required using ADD NEW.
New items can be added at any time.
7. To delete an item, touch the item to highlight it, then
press DELETE ITEM.
8. To clear the item initial weight press CLEAR INITIAL
WEIGHT. This feature is disabled if no initial weight is
stored.
9. To clear the item's final weight, press CLEAR FINAL
WEIGHT. This feature is disabled if no final weight is
stored.
10. To clear all items, press CLEAR ALL, then confirm.
11. When finished, press RETURN TO APPLICATION.
12. On the main Application screen, follow the screen
instructions. Add the first item, then press ACCEPT to
store the value.
13. The initial weights of the defined item are recorded
first, then the final weight so the items are collected
next.
14. Once the item's initial weight and the final weight
are stored, the values are displayed along with the
difference in weight and percent.
15. The items are stored in memory. To begin a new
Differential Weighing session, delete the initial and/
or final weights.
0721 IH-4737