Measuring The Density Of A Liquid Using Acalibrated Sinker (Not Supplied); Measuring The Density Of Porous Material Using Oil - Uline OHAUS EXPLORER H-4737 Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

8. Begin the density determination process as per
above.
MEASURING THE DENSITY OF A LIQUID USING A
CALIBRATED SINKER (NOT SUPPLIED)
1. To enable this feature, enter the density setup menu
and select the following: Density type: Liquid.
NOTE: When the density type is set to liquid, the
liquid type and porous material selections are
disabled.
2. Confirm that the default value displayed (sinker
volume) is correct.
3. To edit the default values, touch EDIT SETTINGS and
the Settings screen is now displayed:
• Settings available: Sinker volume
• Functions available: Return to application
4. To adjust the sinker volume value, touch the SINKER
VOLUME button.
5. A numeric input window is displayed. Key in the
desired value, then press √.
6. The display returns to the previous screen with the
new value highlighted.
7. To return to the density determination home screen,
touch RETURN TO APPLICATION.
8. Press BEGIN DENSITY DETERMINATION to start the
process.
STEP 1 OF 2 – WEIGH THE SINKER IN AIR.
1.
Follow the screen instructions, then press ACCEPT
to store the sinker weight ("in air").
PAGE 23 OF 96
OPERATION CONTINUED
STEP 2 OF 2 – WEIGH THE SINKER SUBMERGED
IN THE TEST LIQUID.
1.
2. Once the necessary weights have been determined,
3. Press BEGIN DENSITY DETERMINATION to reset all the

MEASURING THE DENSITY OF POROUS MATERIAL USING OIL

1. To enable this feature, enter the density
• Density type: Solid
• Liquid type: Water
• Porous material: ON
2. Confirm the default values displayed (water
3. To edit the default values, touch EDIT SETTINGS. The
• Settings available: Water Temp, Oil Density
• Functions available: Return to Application
Follow the screen instructions, then press ACCEPT to
store the sinker weight (submerged in liquid).
the density of the liquid sample is displayed in g/cm
(along with the weight in air, weight in water) on the
application screen.
weight values and restart the process.
determination setup menu, and set the following:
temperature) are correct.
settings screen appears.
3
0721 IH-4737

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ohaus explorer h-4738Ohaus explorer h-4739

Tabla de contenido