Output level
High
Low
High
Low
² "Source Hi_ALM / Source Mi_ALM / Source Lo_ALM": three kinds of alarm sources can trig the nurse
call: high level alarm, medium level alarm and low level alarm (multi-optional). After selecting the
responding alarm level, the device will send out the nurse call signal according to "Source" and "Output
level". If you don't select any source, nurse call signal will not be generated (Note: the item "Source" can
be multi-selected.)
Note:
Nurse Call function can not be regarded as main alarm notice method, please do not entirely rely
on it. You should combine parameter values with alarm level and patient's clinical behavior and symptom
to determine patient's condition.

4.9.7 Network Setup

² "Client IP Address": to set the local IP address for this device working as client.
² "Server IP Address": to set the IP address of remote server (work station) when connecting to a central
monitoring system.
² "Port": the remote port number to which the monitor will connect to the work station in the central
monitoring system. Its setting range is from 6001 to 6064. It can also be used to represent the patient
bed number connecting to the work station. For example, the port number 6002 means the monitor is
assigned to the bed number 2 in the CSM. The work station can connect to up to 64 bedside monitors,
so please set the port between 6001and 6064. Press "
² "HL7 (Protocol)": enable or disable the monitor to export data to CIS/HIS by HL7 protocol. Set it as "On"
means enable HL7 protocol. The default is "On", that's to say, enabled HL7 protocol. When the HL7
protocol is set as "On", then item "Server IP Address" and "Port" are nonadjustable.
User Manual for Vital Signs Monitor
Duration
Continuous
Pulse
Continuous
Pulse
Figure 4.19 Network Setup Screen
43
Output (format)
" to make the new setting effective.