infrared heater, and direct sunlight etc.
Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy.
²
Please do not use the SpO
²
Always observe the plethysmogram (waveform), which is auto-scaled within the range of 100. The
²
SpO
reading may be unlikely true when the waveform is not smooth or irregular. If in doubt, rely on
2
your clinical judgment, rather than the monitor readout.
A functional tester cannot be used to assess the accuracy of the pulse oximeter monitor or a SpO
²
sensor. However, a functional tester, such as SpO
particular pulse oximeter is reproducing the given calibration curve. Before testing the oximeter by a
functional tester, please firstly ask the manufacturer which calibration curve is used, if necessary,
request the manufacturer for its dedicated calibration curve and download it into the tester.
User Manual for Vital Signs Monitor
sensor when having the MRI, or burn may be caused by faradism.
2
61
simulator can be used to check how accurately a
2
2