the tone frequency of pulse beep (sound becomes sharper); The lower the SpO
tone frequency of pulse beep (sound becomes flatter).
² "Color": to set the displaying color for each parameter (such as for SpO
Blue", "Green", "Red", "Light blue" and "Pink" for optional.
² "Date/Time": to set the system date and time, date format: yyyy-mm-dd; time format: hh-mm.
² "Default": to reset to default setting, press "Yes" to reset to the default setting, and press "No" to return
to upper level screen.
² "Touch Calibrate": to enter into touchscreen calibration screen. There will is cross cursor " " appears
on the screen, tap the cross point "
the calibration. Note: after calibration, the monitor will be auto shutdown and auto restart.
² "Print Mode": to set the printing time for real-time printing mode with the options of "Continuous",
"10s", "20s", "30s" and "60s" for optional. "Continuous" means that the device will not stop printing the
real-time plethysmogram and ECG waveform until the user change the display screen or press the print
key again.
XXs: prints the real-time plethysmogram and ECG waveform lasting for XX seconds.
² "About": to enter into system information screen which displays the software version and serial number.

4.10 Alarm Settings

Pressing alarm silence "
There are 3 status in total:
² Alarm sound is enabled, this is the default status..
² Alarm silence lasts for 120 seconds: short pressing alarm silence key, the red icon "
the lower screen, and message "silence count-down time 120". At this time, the alarm silence indicator on
the left side of alarm silence key will be light. The device will mute the alarm sound temporarily for 2
minutes, but keep the visual alarm (lamp) flashing. When the counting down time (120s) is out, the alarm
silence will be de-activated automatically, the red icon "
indicator will be dark.
² Alarm silence lasts all the time: long pressing alarm silence key, the red icon "
lower column of the screen. At this time, the alarm silence indicator on the left side of alarm silence key will
be light. The device will mute the alarm sound in the future, but keep the visual alarm (lamp) flashing. Till a
new type of alarm event is detected, the alarm silence status will be terminated automatically and the alarm
sound will resume, the red icon "
Note: If the current status is alarm silence, longtime or short time pressing the alarm silence key can de-activate
the alarm silence function.
User Manual for Vital Signs Monitor
" of the cross cursor with the stylus or finger one by one to finish
" key to set the alarm sound status.
" will disappear as well, and the alarm silence indicator will be dark.
" will disappear as well, and the alarm silence
45