Bosch DCI 700 Manual Original página 47

Common-rail injector test bench
Tabla de contenido

Publicidad

4. Lubricate the O-ring on the inlet filter with test oil.
Do not set the inlet filter down again. Instead, screw
it into the measurement channel immediately.
5. Install the O-Ring (12 x 2) (1). Only required on inlet
filters with a groove for an O-ring on their front. If
there is no groove for an O-ring, no O-ring is used
during installation.
6. Screw the inlet filter into the measurement channel
and tighten by turning it counterclockwise. Do not
tighten the inlet filter with a tool.
A
B
 48:
Installing the inlet filter
7. Connect the hose assembly
(1)
coupling
on the inlet filter.
A
B
1
2
 49:
Connecting the hose assembly
---Separator---
Robert Bosch GmbH
A
B
1
(2)
to the quick-release
A
B
4.2.5
Changing the protective cover
1. End all computer applications.
2. Shut down the Windows operating system.
3. Switch off the DCI 700 at the main switch (4).
4. Open the protective door
(2)
door contact
for the safety switch must be loca‐
ted outside the switch box for door monitoring (1).
3
1 2
 50:
Preparing to replace the protective cover
5. Remove 10x countersunk bolts
6. Remove the edge protection frame with the protec‐
(2)
tive cover
and place it on a clean surface.
2
1
 51:
Removing the protective cover with edge protection frame
7. Remove 16x hex nuts (1).
(2)
8. Remove 8x mountings
9. Remove the protective cover
DCI 700 | 47 | en
(3)
approx. 15 cm. The
4
(1)
.
.
(3)
.
1 689 989 473 | 2020-11-24

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido