PCB Jumper Positions

Only the analog version of the Main board has a jumper on it. This section provides the details for
this jumper.
2.6.1.

Main PCB Jumper

The IDNet version of the IND570 main board has no jumpers. There is one jumper on the IND570
analog scale main board (W1). This jumper selects either 2 mV/V or 3 mV/V operation of the
analog circuitry. 3mV/V is the factory default position. Refer to Figure 2-61 for the jumper location
and Figure 2-62 for its settings.
The 2 mV/3 mV jumper settings are described in Table 2-7.
2-42
METTLER TOLEDO IND570 Weighing Terminal Installation Manual
Dis
Figure 2-61: W1 Millivolt Jumper Location, Main PCB
When removing the W1 jumper for 3mV/V operation, reposition it on just one of the pins, as
shown in Figure 2-61 and at left in Figure 2-62. If this jumper is not correctly positioned, the
plastic cover over the Main board will not fit properly. If this occurs, do not force the plastic
cover down. Remove the cover and position the jumper as shown.
Figure 2-62: W1 Millivolt Jumper Positions
Jumper
Setting
OPEN
W1
CLOSED
SW1
Table 2-7: Millivolt Jumper Description
Description
Sets the connected scale's load cell setting to 3 mV/V
Sets the connected scale's load cell setting to 2 mV/V.
30205319
|
04
|
08/2016