Clean any oil or other contaminants from the area of the overlay shown in Figure 2-65 where the
capacity label will be added. Peel the backing from the label and adhere it to the overlay in the
location shown in Figure 2-65, or another location acceptable to the local regulations.
2.7.2.1.
Special Tare Marking for Multi-Interval Weighing Applications
Because of the way the IND570 manages the tare function in multi-interval applications (refer to
section 2.7.2.3 in Operation), it is required that the final weighing system has a label displayed
that indicates the maximum tare value within the capacity of interval # 1: T = nnn, where nnn is the
capacity of interval 1.
The requirements for this marking are as follows:
2-44
METTLER TOLEDO IND570 Weighing Terminal Installation Manual
Figure 2-64: Preparing the Capacity Label
Capacity Label
in Place
Required only for OIML approved terminals that are programmed for multi-interval (not multiple
range).
The descriptive markings shall be indelible and of a size, shape and clarity allowing easy
reading.
It shall be located in a clearly visible place on a sticker fixed permanently to the instrument.
In case the sticker is not destroyed when removed, a means of securing shall be provided, e.g.
a control mark that can be applied.
Figure 2-65: Capacity Label Installed
Cut at dotted line to remove
unused ranges/intervals
30205319
|