2.4.5.3.
IDNet Connections
The IND570 terminal supplies 12 V for the newer T-Brick type cell of the IDNet base. Older types of
IDNet bases (known as Pik or Pik-Brick) require both a 12 V and a 32 V supply. IND570 does not
support older IDNet bases that require a 32 V supply.
When using an IDNet version of the IND570 terminal, the cable connection from the base is made
to a connector (Figure 2-29) on the rear of the housing. The IDNet bases are supplied with a length
of cable and a connector that mates to the connector on the IND570 terminal.
30205319
|
04
|
08/2016
Note for the standard four-wire cable: If an increase in load results in a decrease in weight
display, reverse the signal wires (+SIG and −SIG).
Figure 2-29: IDNet Connector Location on the Harsh (left) and Panel (right) Enclosures
METTLER TOLEDO IND570 Weighing Terminal Installation Manual
Figure 2-28: Load Cell Termination
2-19