2.8.2.
External Sealing of the Harsh Enclosure
For external sealing of the harsh enclosure, refer to Figure 2-66 and Figure 2-70 and follow these
steps:
1. Ensure that the appropriate approval region has been selected in setup under Scale > Type >
2. With the front panel installed on the enclosure and snapped into place, thread the free end of
3. Thread the end of the wire cable through the hole in the plastic seal (as shown in Figure 2-70),
2-46
METTLER TOLEDO IND570 Weighing Terminal Installation Manual
Figure 2-68: External Sealing for United States
Figure 2-69: External Sealing for Canada
Approval and that the Metrology security switch SW1-1 is in the "on" position.
Refer to section 2.1.2.2, Closing the Harsh Enclosure, for the correct method to use when
replacing the front panel.
the wire seal through either the left or right hole in the IND570 front panel, and through the hole
in the retaining clip.
remove any remaining slack in the wire, and snap the seal shut.
30205319
|
04
|
08/2016