Use the MAQUET TEGRIS interface
The MAQUET TEGRIS interface 15 is used to connect the device to the
MAQUET TEGRIS system for central device control.
When the device is connected, all the main functions and?settings of the
camera system:
can be controlled via the TEGRIS control and
they are displayed on the TEGRIS screen.
When the MAQUET TEGRIS system is connected and active, all the func-
tions and settings of the camera system can be set and displayed via the
device and via the MAQUET TEGRIS system.
If both are being operated at the same time, entries on the device take
precedence. Doctors and the nursing or service personnel must be thor-
oughly familiar with how to set and operate the MAQUET TEGRIS system,
see TEGRIS documentation for information on settings, use and trouble-
shooting.
Connecting the power supply
Risk of death by electric shock!
Connect the product only to a grounded power
supply.
DANGER
Insert the power cord on the rear panel of the camera control unit into
the mains connector 6.
Plug in the mains plug at a socket of the building mains.
Plug in the equipotentialization cable into the equipotentialization
connector 10 of the camera control unit.
Connect the plug of the equipotentialization cable to the building
mains equipotentialization connector.
Make certain that the rear of the device is accessible at all times.
5.2

Function checks

Camera control unit
Check camera control unit for external damage and signs of impact or
other mechanical impacts.
Power cord
Check that the power cord is not broken, kinked or twisted.
Video cable
Check that the video cable is not broken, kinked or twisted.
8
®
Camera head with endoscopic lens and camera cable
Note
The endoscopic lens is firmly attached to the camera head and therefore
cannot be removed!
Risk of infection due to soiled or contaminated
products!
WARNING
Disinfect camera head and camera cable before
each use, or use with sterile cover.
Check that the front glass of endocoupler on the camera head is clean.
If necessary, clean the glass with isopropyl alcohol (70 %).
Make certain that adjusting ring for the focusing ring 18 turns freely
without sticking.
Check that camera cable 26 is not broken, kinked or twisted. Do not
use the camera head in the event of damage.
Check that pins in connection plug 25 are not bent and that there is
no moisture in the plug.
Ensure that the housing of the camera head is not damaged or broken.
Do not use the camera head in the event of damage.
Aim camera head to an object in the room, focus the image and check
for good image quality.
Sterile cover
Note
Aesculap recommends using disposable sterile covers, (e.g. art. no. JG904).
Use of the sterile covers is described in the relevant instructions for use.
Check that the packaging of the sterile covers is unopened, undamaged
and dry before using the sterile covers in the operating room.
5.3

Safe operation

Risk of injury and/or malfunction!
Always carry out a function check prior to using
the product.
WARNING
Risk of injury when using the product beyond the
field of view!
Apply the product only under visual control.
WARNING
Note
The camera control unit requires approx. 30 s to boot up.