3. variant: Connecting the camera and monitor with several documen-
tation devices daisy-chained between them (for parallel video images
on the monitor)
If video images from the documentation devices are intended to be
overlaid on the monitor display, connect Device 1 and Device 2 (e.g.
digital documentation system) between the camera and Monitor 1.
Select signal or type of video cable according to the common signal
standard of the camera and monitor, always choosing the highest-
quality signal option (video signals in descending order of quality: DVI,
HD-SDI, S-Video).
Connect Monitor 2 to Monitor 1 if the video images from the docu-
mentation devices are to be displayed on both monitors.
If necessary, connect remote cable.
Note
The optional direct video connection between the camera and the monitor
is recommended as an additional signal line if the video signal can be
looped through device 1 and device 2 only with the devices switched on
(see instructions for use of the devices). This ensures that, in case of a fail-
ure of Device 1 or Device 2, the monitor can be immediately switched to the
direct video signal.
The alternative direct video connection between Monitor 2 and the camera
is recommended if Monitor 2 is only to display the live image.
Stacking of units
Make certain the system is set up on a sufficiently stable support (e.g.
a table, ceiling support, equipment cart, etc.).
Do not exceed a stack height of 300 mm.
Moving stacks of units
Always lift stacks from underneath the bottommost unit.
Presetting the correct voltage
The mains voltage must correspond to the voltage indicated on the type
plate at the back of the unit.
5.
Working with the camera control unit
Full HD PV470 and the camera head
PV471/PV472/PV473
5.1

System set-up

Connecting the accessories
Risk of injury due to unapproved configuration
using additional components!
DANGER
For all applied components, ensure that their
classification matches that of the application
component (Type CF) of the respective device.
Risk of injury when using devices or components
that are not of type CF!
Do not touch the device and patient at the same
DANGER
time.
Do not place the patient in direct contact with
the device.
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
Also note that any equipment connected at the interfaces must demon-
strably meet the respective IEC standards (e.g. IEC 60950 for data process-
ing equipment, IEC/DIN EN 60601-1 for electromedical devices).
All configurations must comply with basic standard IEC/DIN EN 60601-1.
Any individual connecting devices with one another is responsible for such
configuration and must ensure compliance with basic standard IEC/DIN
EN 60601-1 or applicable national standards.
Please address your B. Braun/Aesculap partner or Aesculap Technical
Service with any inquiries in this respect; for a contact address, see
Technical Service.
Connect video cable
Connect video cables according to diagrams, see First use.
Communication interface connection (optional accessory)
With the communication interface connection, the LED light source
OP940 can be switched on and to standby via the FULL HD CMOS camera
system - see connection diagrams 1, 2 or 3. In order to use this remote
control of the light source, the communication interface cable OP942
must be connected to the two devices.
Connect the communication interface cable OP942 between the LED
light source OP940 and the camera control unit 11.