4.

Preparation and setup

Non-compliance with the following instructions will preclude all respon-
sibility and liability in this respect on the part of Aesculap.
Danger of explosion from a device that is improp-
erly set up!
Make certain that the power plug is connected
DANGER
to the power supply outside of any areas where
there is a risk of explosion.
Have ready replacement products/camera sys-
tem.
When setting up and operating the product, adhere to
– national regulations for installation and operation,
– national regulations on fire and explosion protection.
Note
For the safety of patients and users it is essential that the mains power cord
and, especially, the protective earth connection are intact. In many cases
defective or missing protective earth connections are not registered imme-
diately.
Note
The potential equalization plugs of the device must be connected to the
potential equalization band (see IEC/DIN 60601-1-1 or corresponding
national standards.)
Note
The connection of electrical devices to a multiple socket creates a system
and can lead to a reduced level of safety.
Note
If the equipment cart is set up outside of the sterile area, measure the cable
length to give enough room for movement so that the sterility is not com-
promised.
Connect the device via the potential equalization terminal at the rear
panel of the device to the potential equalization system of the room
used for medical purposes.
Ensure that all devices operated in the vicinity meet their relevant EMC
requirements.
Note
The potential equalization lead can be ordered from the manufacturer as
art. no. GK535 (4 m length) or TA008205 (0.8 m length).
6
®
4.1

Preparation

Non-compliance with the following instructions will preclude all respon-
sibility and liability in this respect on the part of Aesculap.
First use
Risk of injury and/or product malfunction due to
incorrect operation of the electromedical system!
WARNING
Do not use the device in areas where there is a
risk of explosion or where there is a high level
of oxygen.
Adhere to the instructions for use of any medi-
cal device.
Do not use the device in a magnetic resonance
environment.
There are three different connection variants, depending on system
requirements.
If a parallel video image from the documentation device is not
required, always connect the components as shown under variant 1,
see fold-out page.
If a parallel video image from the documentation device is required on
a monitor, connect the components as shown under variant 2 or 3, see
fold-out page.
Always connect the video output (Video out) of one device to the video
input (Video in) of the next device.
1. variant: Connecting the camera, monitor and one documentation
device to the camera control unit (no parallel video images)
Select signal or type of video cable according to the common signal
standard of the camera and monitor, always choosing the highest-
quality signal option (video signals in descending order of quality: DVI,
HD-SDI, S-Video).
If necessary, connect remote cable.
2. variant: Connecting the camera and monitor with one documenta-
tion device daisy-chained between them (for parallel video images)
If video images from the documentation device are intended to be
overlaid on the monitor display, connect the device (e.g. digital docu-
mentation system) between the camera and the monitor.
Select signal or type of video cable according to the common signal
standard of the camera and monitor, always choosing the highest-
quality signal option (video signals in descending order of quality: DVI,
HD-SDI, S-Video).
If necessary, connect remote cable.
Note
The optional direct video connection between the camera and the monitor
is recommended as an additional signal line if the video signal can be
looped through the documentation device only with the device switched on
(see instructions for use provided by the manufacturer of the respective
device). This ensures that, in case of a documentation device failure, the
monitor can be immediately switched to the direct video signal.