Sterilization; Inspection, Maintenance And Checks; Packaging; Storage - Braun Aesculap PV470 Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 64
Phase I
Fully immerse the product for at least 5 min in the cleaning solution.
Ensure that all accessible surfaces are moistened.
Clean the product immersed in the solution, using a soft cloth or, if
necessary, a soft cleaning brush, until all visible residues have been
removed from the surfaces.
If applicable brush through all surfaces that are not accessible to visual
inspection for at least 1 minute.
During cleaning, mobilize non-rigid components, such as set screws,
joints, slides, etc., 3 times in each direction to the positive stop.
After cleaning, use a suitable disposable syringe to thoroughly rinse
these parts of the product that are difficult to access at least 3 times.
Phase II
Rinse/flush the product (all accessible surfaces) thoroughly for at least
1 minute under running water.
During rinsing, mobilize non-rigid components, such as set screws,
joints, slides, etc., 3 times in each direction to the positive stop.
Perform these steps at least 3 times.
Drain any remaining water fully.
Phase III
Fully immerse the product for at least 12 minutes in the disinfectant
solution.
During rinsing, mobilize non-rigid components, such as set screws,
joints, slides, etc., 3 times in each direction to the positive stop.
Rinse lumens at least 3 times at the beginning of the exposure time,
using a disposable syringe. Ensure that all accessible surfaces are
moistened.
Phase IV
Following disinfection, rinse/flush the product completely (all accessi-
ble surfaces) for at least 2 minutes.
During rinsing, mobilize non-rigid components, such as set screws,
joints, slides, etc., 3 times in each direction to the positive stop.
Rinse lumens at least 3 times at the beginning of the exposure time,
using a disposable syringe.
Perform these steps at least 3 times.
Drain any remaining water fully.
Phase V
Dry the product with suitable tools.
6.7

Sterilization

Damage to the camera control unit by sterilization!
Do not sterilize the camera control unit under
any circumstances.
CAUTION
Clean the camera control unit by surface disin-
fection only.
Damage to the camera head and the adapter
caused by inappropriate sterilization processes!
Never sterilize the camera head and the adapter
CAUTION
with steam or in an autoclave.
Do not carry out chemical sterilization.
Working with disposable sterile covers
Use suitable Aesculap disposable sterile covers. Follow the instructions
for use of the sterile covers.
Approval of compatibility of materials exists for:
ETO-sterilization
6.8

Inspection, maintenance and checks

Allow the product to cool down to room temperature.
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
Set aside the product if it is damaged.
6.9

Packaging

Follow the instructions for use for the applied packaging and storage
systems (e.g. instructions for use TA009721 for Aesculap Eccos® stor-
age system).
Insert the product in its proper position in the Eccos® holder, or put it
on a tray in such a way that the product is protected against damage.
Ensure that all cutting edges are protected.
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
Ensure that the packaging will prevent a recontamination of the prod-
uct.

6.10 Storage

Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
Store sterile packed single-use products dust-protected in a dry, dark
and temperature-controlled room.