Table 3 – Test levels for Radiated and conducted Immunity Tests
Guidance and manufacturer's declaration – electromagnetic immunity
The IMAGE1 S™ is intended for use in the electromagnetic environment specified below. The user of IMAGE1 S™ should make sure that it is used in
such an environment.
Immunity tests
EN/IEC 60601 test level
Conducted RF
3 V
on 150 kHz to 80 MHz
rms
disturbances
6 V
in ISM bands between
rms
acc. to IEC 61000-4-6
0.15 MHz and 80 MHz
Radiated RF
3 V/m
disturbances acc. to
80 MHz to 2.7 GHz
IEC 61000-4-3
Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected by absorptions and reflections of buildings,
objects, and persons.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the IMAGE1 S™ is used exceeds the applicable
RF compliance level above, the IMAGE1 S™ should be observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as re-orienting or relocating the IMAGE1 S™.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Information on
electromagnetic
compatibility (EMC)
Compliance level
Electromagnetic environment – guidance
Portable and mobile RF communications equipment should be
used no closer to any part of the IMAGE1 S™, including cables,
than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance:
f
3 V
on 150 kHz to 80 MHz
rms
6 V
in ISM bands between
rms
where P is the nominal power of the transmitter in watts [W]
0.15 MHz and 80 MHz
according to the information provided by the manufacturer of the
transmitter and d is the recommended separation distance in
meters [m].
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
level in each frequency range
f
3 V/m
80 MHz to 2.7 GHz
h
Interference may occur in the vicinity of equipment marked with
the following symbol:
7
Indicaciones sobre
compatibilidad
electromagnética (CEM)
should be less than the compliance
a
b
.
80 MHz to 800 MHz