Recommended separation distances between portable and mobile RF communications equipment and the TELE PACK VET X model 69 0450 20
The TELE PACK VET X model 69 0450 20 is intended for use in an electromagnetic environment in which the RF disturbances are controlled. The user of the
device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (trans-
mitters) and the device, according to the output power of the communications equipment as recommended below.
150 kHz to < 80 MHz
Nominal power of trans-
mitter
3
d
=
[W]
0.01
0.12
0.1
0.38
1
10
100
For transmitters whose maximum nominal power is not listed in the above table, the recommended safety distance d in meters [m] can be determined using
the equation belonging to the relevant column, where P is the maximum nominal power of the transmitter in Watt [W] as specified by the manufacturer of the
transmitter.
Note: These guidelines may not apply in all situations. The propagation of electromagnetic quantities is affected by absorptions and reflections of buildings,