en
4.
Product description
4.1
Scope of Supply
Art. no.
Designation
GA666
Acculan NiMH battery short
– or –
GA676
Acculan NiMH battery
TA011867
Instructions for use, Acculan battery GA666, GA676
4.2
Components required to operate the device
■
Acculan 4 charger GA320 or Acculan 3Ti charger GA677
■
Acculan 4 or Acculan 3Ti machine:
Art. no.
Designation
GA330
Acculan 4 drill and reamer
GA331
Acculan 4 oscillating saw
GA334
Acculan 4 reciprocating saw
GA336
Acculan 4 oscillating saw straight
GA337
Acculan 4 oscillating sternum saw
GA340
Acculan 4 Dermatome
GA341
Acculan 4 Dermatome 0.1 mm
GA344
Acculan 4 small drill
GA668
Acculan 3Ti oscillating saw straight
GA670
Acculan 3Ti Dermatome
GA671
Acculan 3Ti small drill
GA672
Acculan 3Ti drill and reamer
GA673
Acculan 3Ti oscillating saw
GA674
Acculan 3Ti reciprocating saw
GT666
Acculan 4 TPLO saw
GT668
Acculan 3Ti TPLO saw
4.3
Operating principle
The Acculan NiMH battery supplies the Acculan machine with voltage. The
battery is charged in the Acculan 4 charger GA320 or Acculan 3Ti charger
GA677. The charging process of a completely discharged battery can take
up to 120 min.
The speed and rotational direction are set depending on the trigger setting
of the machines:
■
for saws and Dermatome: speed clockwise
■
for drills: speed clockwise, counter-clockwise, oscillation and pilgrim
step
5.
Preparation
►
Prior to use, inspect the product and its accessories for any visible dam-
age.
►
Use the products and their accessories only if they are in perfect tech-
nical condition.
4
6.
Working with the device
6.1
System set-up
►
See instructions for use for the respective Acculan machine and the
Acculan 4 charger GA320 or Acculan 3Ti charger GA677.
6.2
Functional test
See the instructions for use for the respective Acculan machine.
6.3
Safe operation
WARNING
Risk of injury and material damage due to inappropriate use of bat-
teries!
►
Charge battery only in the Acculan 4 charger GA320 or
Acculan 3Ti charger GA677.
►
Insert battery correctly (Acculan machine and charger).
WARNING
Risk of injury and material damage due to damaged batteries!
►
Check batteries for any damage before charging.
►
Do not use or recharge damaged batteries.
CAUTION
Total battery failure can occur if the batteries are left in the charger
while it is disconnected from the power supply!
►
Always keep the charger connected to the power supply. Indicator
light for readiness for service must be lit.
►
Always remove batteries from the charger when it is disconnected
from the power supply.
CAUTION
Loss of capacity/performance due to prolonged storage of discharged
batteries!
►
Leave the batteries in the functional charger after they have been
successfully charged (trickle charging).
►
For proper loading and removal of the battery from the Acculan
machine, see the instructions for use for the respective Acculan
machine.
7.
Processing process
7.1
General safety information
Note
Observe national statutory regulations, national and international stan-
dards and directives, and any proprietary hygiene instructions for process-
ing.
Note
The applicable national regulations concerning the reprocessing of devices
must be observed for patients having Creutzfeldt-Jakob disease (CJD), sus-
pected of having CJD, or possible variants of CJD.
Note
It should be noted that successful processing of this medical device can only
be guaranteed following prior validation of the processing method. The
operator/processor shall be responsible.