5.4.1.2
Edit: Source
Procedure
1.
Press the Utility key.
2.
Press the More key.
3.
Press the Go-NoGo Menu key.
4.
Press Source repeatedly to select the source channel (CH1 or CH2).
5.4.1.3
Edit: NoGo Violation Conditions
Procedure
1.
Press the Utility key.
2.
Press the More key.
3.
Press the Go-NoGo Menu key.
4.
Press Violating repeatedly to select the NoGo conditions.
5.4.1.4
Edit: Template (boundary)
The NoGo template sets the upper and lower amplitude boundary. Two
methods are available: Min/Max and Auto.
Min/Max
11-2011
Selects the upper boundary (Max) and lower boundary (Min)
as separate waveforms, from the internal memory. The
upper boundary is saved to Ref A, the lower boundary is
saved to Ref. B.
Advantage: The template shape and distance (allowance)
between the source signal are fully customizable.
Disadvantage: The waveforms (templates) have to be
stored internally prior to this selection.
USER'S MANUAL. OD-405/410B/411/415