3.
Press ProbeComp.
4.
Press Wavetype repeatedly to select the standard square wave.
5.
Press the Autoset key. The compensation signal will appear in the
display.
6.
Press the Display key, then Type to select the vector waveform.
7.
Turn the adjustment point on the probe until the signal edge
becomes sharp.
11-2011
USER'S MANUAL. OD-405/410B/411/415
Figure 28.-
Figure 29.-
Page 98