USER'S MANUAL. OD-405/410B/411/415
Auto
Creates the upper and lower boundary (template) from the
source signal, not from an internally stored waveform.
Advantage: No need to store the waveforms prior to this
selection.
Disadvantage: The template shape is proportional to the
source signal. The distance (allowance) between the source
signal and the upper and lower template is the same.
Max/Min
1.
The template is based on the source signal. Ensure the source
signal appears on the display.
2.
Press the Utility key.
3.
Press the More key.
4.
Press the Go-NoGo Menu key.
5.
Press the Template Edit key.
6.
Press Template repeatedly to select the upper (Max) or lower (Min)
boundaries.
7.
Press Source and use the Variable knob to select the waveform
template.
Max Waveform A: Ref A, W01~W15
Min Waveform B: Ref B, W01~W15
8.
Press Position and use the Variable knob to set the waveform
amplitude.
9.
Repeat steps 5-7 for the other template setting (Max or Min).
10. When both Max and Min templates have been configured, press
Save & Create to save the templates.
Page 51
11-2011