Vercise™ Deep Brain Stimulation Systems
6. With the Lead Stylet in place, slide the OR Cable Connector onto the proximal end of the DBS
Lead (Figure 17). Make sure that the DBS Lead is fully inserted.
Note: The Lead Stylet will extend through the back hole of the OR Cable Connector when the
DBS Lead is inserted as shown in Figure 17.
7. Hold the DBS Lead in place. Slide the locking lever to the locked (1) position (Figure 17).
Figure 17. Securing the DBS Lead Into the 1x8 OR Cable Connector
8. Support the OR Cable Connector to prevent unnecessary bending of the Lead during testing.
9. Verify that impedances are acceptable by using the Clinician Programmer or Remote Control.
10. Evaluate Lead placement by the appropriate methods. If necessary, adjust the Lead location or
stimulation parameters.
Note: The stylet should remain in place throughout insertion or adjustments of the DBS Lead.
High charge density can cause permanent tissue damage. The Clinician
Warning:
Programmer will limit the stimulation parameters to safe values.
Warning:
Increasing the number of Lead penetrations increases the probability of
hemorrhage. To minimize acute Lead revisions, use techniques of target
localization, such as microelectrode recordings and/or imaging.
11. Turn off ETS 2.
A sudden increase in stimulation may occur if ETS 2 is ON while disconnecting
Warning:
the OR Cables.
12. Slide the locking lever to the open (0) position. Disconnect the OR Cable Connector and
Extension from the proximal end of the DBS Lead.
13. Verify that the DBS Lead has not moved from the desired location.
Surgical Implant Manual
92495783-02
Page 32 of 1157