IMPORTANT: If the mute button is pressed prior to an alarm condition (for example, to
mute the device in a movie theatre), critical high priority and high priority alarms will
override the mute function; medium and low priority alarms will be muted for eight hours
and twenty four hours respectively from the time the button was pressed. Press the mute
button off to display the last highest priority alarm. Press the mute button on again to
reset the eight-hour timer.

9 . 1 . ALARMS

EN
When the concentrator sounds an alarm, a corresponding message will be displayed on
the screen. Take appropriate action as directed in the charts below.
9 . 1 . 1 . Critical High Priority Alarms
IMPORTANT: These alarms will disable the device immediately.
ALARM
MESSAGE
*Value: 01-20
9 . 1 .2 . High Priority Alarms
IMPORTANT: These alarms will allow the device to continue operating.
ALARM
MESSAGE
DESCRIPTION
Battery needs charging.
Battery is not an approved battery
Service required.
DESCRIPTION
Device is unable to maintain oxy-
gen purity.
Estimated battery life less than 17
minutes.
Recharge the battery pack by plug-
ging in to the power supply. Ensure
all connections are made securely.
Replace battery with an approved
battery.
Contact your distributor.
Be sure air inlet/outlet has not
been blocked. If alarm persists,
contact your distributor.
Charge the battery pack by plug-
ging in to power supply.
Important: The message will be au-
tomatically cleared when plugged
in to power supply.