15 .3 . EU RED ARTICLE 10(10) RESTRICTIONS ON USE FOR MODEL RS-00500C:
Based on the information available from the European Communication Office (ECO) Fre-
quency Information System (EFIS), Zen-O™ (model RS-00500C) complies with stipulated
radio frequency radiation limits. The table below demonstrates compliance to Article
10 (10) of the European Union Radio Equipment Directive (RED) and confirms that there
are no restrictions with regards to its use by patients in member states of the European
Union;
RF Transmitter Information:
UPLINK /
FUNCTION
/ BAND US-
TRANSMIT
AGE
(MHZ)
EGSM 900
880.0 – 915.0
DCS 1800
1710.2 – 1784.8
GPS
N/A
GLONASS
N/A
GALILEO
N/A
BTLE
2400 – 2483
15 .4 . RF EXPOSURE INFORMATION:
The RS-00500C model meets the applicable limits for radio frequency (RF)exposure, as
determined by independent tests for the Specific Absorption Rate (SAR).
Specific Absorption Rate (SAR) refers to the rate at which the body absorbs RF energy.
The SAR limit is 1.6 watts for kilogram in countries that set the limit averaged over 1 gram
of tissue and 2.0 watts per kilogram in countries that set the limit averaged over 10 grams
of tissue.
The highest SAR values measured in accordance with ICNIRP guidelines for RS-00500C
are:
Body SAR at 1g: 1.50 W/Kg
Body SAR at 10g: 1.70 W/Kg
During use the actual SAR values for this device are usually well below the values stated
above, because for purposes of system efficiency operating power is reduced from full
when not needed and the lower the power output the lower its SAR value.
DOWNLINK /
MODULATION
RECEIVE
SCHEME
(MHZ)
925.0 – 960.0
GMSK
1805.2 – 1879.8 GMSK
1563 – 1587
CDMA / FDMA
1593 – 1610
CDMA / FDMA
1559-1610MHZ
2400 – 2483
GFSK/Π/4-
DPSK/ 8-DPSK
39/338
MAXIMUM EFFEC-
TIVE RADIATED
POWER
(ERP/ EIRP)
+9.33DBM / 8.57MW
+18.79DBM / 75.68MW
RECEIVE ONLY
RECEIVE ONLY
/
RECEIVE ONLY
-8.0DBM
(EIRP)
EN
/0.158MW