10 . TROUBLESHOOTING

PROBLEM
System becomes
inoperative
Any alarm sound or
either
or
LED lit
Battery not charging
POSSIBLE CAUSE
• System may be dis-
connected from the
power source.
• System may be
turned off.
• Critical high priority
alarm has occurred.
• See Chapter 9 -
Alarm Indicators.
• Power is not
connected.
• Battery is not fully
inserted.
• Battery is inoperable.
29/338
TROUBLESHOOTING
• Check that the system is connected
securely to the power source.
• Ensure the system is powered on.
• Examine the system for damage or
exposure to liquids.
• If problem persists, contact your
distributor.
See Chapter 9 - Alarm Indicators.
• Check connections to ensure:
• Round receptacles are secure in unit.
• Power cord is connected to AC/DC
supply or automotive DC adapter is
connected, if applicable.
• Power cord is connected to wall out-
let, if applicable.
• Wall outlet has power.
• Ensure battery is fully seated and bat-
tery cover is secure.
• If problem persists, contact your
distributor.