ADJUSTING AND SECURING GRID PLATFORM
1. Loosen Vertical Adjustment Screws with the provided hex wrench  (3/32"). Align needle with the grid display on ultrasound monitor by turning the
Vertical Adjustment Knob until the vertical needle position is correctly aligned with display.  Tighten Vertical Adjustment Screws.
2. Loosen Horizontal Adjustment Screw with the provided hex wrench (3/32"). Align needle with the grid display on ultrasound monitor by turning the
Horizontal Adjustment Knob until the horizontal needle position is correctly aligned with display. Tighten Horizontal Adjustment Screw.

DEVICE INSPECTION

1. Stepper should hold transducer securely, allow smooth rotation and precise movement along longitudinal axis of the transducer.
FUNCTIONAL PREUSE CHECKS
WARNING
Prior to each use perform the following checks to ensure optimal stepper performance.
n
1. Rotate carriage knob to ensure carriage moves freely through full range of linear travel and performs stepping function.
2. Rotate cradle to ensure movement through full range of rotational travel.
I. Vertical Adjustment Screws
II. Vertical Adjustment Knob
III. Horizontal Adjustment Screw
IV. Horizontal Adjustment Knobs