EX3 Stepper
ADJUSTING AND SECURING GRID PLATFORM
1. Loosen Vertical Adjustment Screws with the provided hex wrench  (3/32"). Align needle with the grid display on ultrasound monitor by turning the
Vertical Adjustment Knob until the vertical needle position is correctly aligned with display.  Tighten Vertical Adjustment Screws.
2. Loosen Horizontal Adjustment Screw with the provided hex wrench (3/32"). Align needle with the grid display on ultrasound monitor by turning the
Horizontal Adjustment Knob until the horizontal needle position is correctly aligned with display. Tighten Horizontal Adjustment Screw.
DISASSEMBLING EX3 STEPPER FOR REPROCESSING
1. Place EX3 Stepper onto flat surface.
2. Loosen hold screws (3) by turning counterclockwise until top cover can be lifted vertically.
3. Rotate stepper, loosen hold screws (2) by turning counterclockwise until bottom cover can be removed. Ensure USB cord is disengaged from cover.
4. Loosen interface screws (2) by turning counterclockwise.
I. Vertical Adjustment Screws
II. Vertical Adjustment Knob
III. Horizontal Adjustment Screw
IV. Horizontal Adjustment Knobs
I. Hold Screws
II. Top Cover
3
English
I. Bottom Cover
II. Hold Screws