Use short, gentle strokes, holding the clipper blade assembly at a flat to slight angle to the patient's
skin (e). For best results, clip against the direction of the hair growth. For contoured areas, stretch
the skin taut while clipping. There are several ways to grasp the clipper. Determine the most
comfortable technique that allows you to keep the clipper blade assembly at a flat to a slight angle
to the skin. This technique can be performed in a push or pull motion. Never rake the clipper over
the skin (f).
e.
To Change Disposable Blade:
Push the ON/OFF button into the OFF position. Hold the clipper body in one hand. Position the
clipper over an appropriate waste receptacle with the blade facing downward. With your thumb,
push the blade release button to push the blade off the clipper body into waste receptacle (g).
NOTE: To reduce cross contamination between patients, remove the clipper blade assembly and
discard appropriately after each use.
Care and Cleaning of the Clipper Body:
The housing of the clipper body is made of durable plastic and is water resistant. It may be wiped off
with a damp cloth containing a small amount of alcohol, disinfectant, or a mild detergent solution. It
can also be rinsed under running water or submersed in a high level disinfectant solution. Approved
chemicals for use in cleaning include 70% alcohol, 2% bleach, Quaternary ammonium chloride,
or Chlor‑clean solution. Allow to dry completely before placing clipper back into drop‑in charger
or reusing.
Do not sterilize the clipper body or blade assembly.
Charging the 9681 Battery‑Operated Clipper:
To charge the clipper:
1. Push the ON/OFF button to the OFF position.
2. Clean the clipper body before returning clipper to the drop‑in charger, making sure the
clipper and charger contacts are clean and free of hair or other contaminants.
3. Connect the cord of the drop‑in charger to an appropriate electrical outlet, making sure the
cord is not pinched or kinked.
4. To be certain that the 9681 clipper is fully charged, plug it in for 4 hours before its initial use
and every use thereafter.
5. We recommend leaving the clipper body in the drop‑in charger in between uses. It cannot
be overcharged.
Storage/ Shelf Life/Disposal
Storage and transport conditions:
‑20°C to 45°C, <75% RH, 86 to 106kPa
Operating conditions:
30% to 75% RH, 70 to 106kPa
0°C to 29°C: no duty cycle
>30°C to 35°C: using a 30‑minute‑on followed by 20‑minute‑off duty cycle