WARNING
If the automatic turn-ON function is set it will be impossible to turn off the
device during the test. The closed lock icon located on the top center of the
screen warns the user of the current setting.
Select the item by touching the display and choose between the following options:
Manual
Automatic
Manual turn-ON: allows to set the turning on of the device by using the specific ON/OFF key.
Automatic turn-on: enables the programming of the frequency and duration the device turns on. Select the desired item and press OK.
If the user selects automatic turn-on she/he may choose from the following options:
One time only
Once a week
Monday to Friday
Saturday-Sunday
Every day
For each option a menu appears to set the day and time that the device will turn on
and off.
If turn ON mode was previously enabled, and the user turns on the device at a
different time from that of the programmed time, the device will show the screen on
the right. To continue touch the OK icon and insert the PIN. If the PIN is not set
insert the default PIN which is the following:
by touching the
the device will turn off.
Oximetry setup
When entering the Oximetry Setting menu the following items are shown:
Alarms Setting
SpO2 Sampling Rate
Pulse tone ON/OFF
Default alarms
Alarms setting
Access to these settings is password protected and allows the user to set the lower and
upper threshold values for SpO
warn the user if the SpO
and / or Pulse Rate values fall below the minimum threshold or
2
exceed the previously set maximum threshold.
The configurable parameter are the lower/upper thresholds of the SpO
parameters. For each parameter is possible to set the alarm ON or OFF and to change the
Default threshold value by the modify icon
The table shows the lower and upper threshold values may be set:
Alarm Limit
SpO
SpO
BPM min
BPM max
WARNING
SPIRODOC
1 2 2 3
and Pulse Rate; during the test, an acoustic alarm will
2
.
Minimum
min
2
max
2
cod. 980156
and Pulse Rate
2
Maximum
85
99
85
99
30
235
30
240
Rev 4.2.1
spirodoc
User manual
EN 26/48