It is possible to select the type of calculated parameters during spirometry and oximetry testing. For each of the two categories the user
may select from the following three options:
simplified
personal
complete
The "simplified" mode only allows for the main parameters to be viewed according to the main standards in effect. In paragraphs 2.7.1.
and 2.7.2 these parameters are shown.
In "personal" mode the user may select which parameters will be displayed. The parameters highlighted in white will be displayed. To
remove a parameter from the list simply touch the white highlighted parameter and it will turn grey.
In "complete" mode at the end of a test all the parameters that the device is capable of calculating will be shown.
WARNING
The parameters of the "simplified" mode are always shown regardless of which mode has been selected.
WARNING
Certain oximetry parameters are grouped together according to the kind of information displayed; by selecting one parameter
from a group all the other parameters belonging to the same group will selected automatically.
WARNING
When the NHAHES III standard is selected the spirometry parameter setup function will be automatically disabled.
Select turbine
Select the option by touching the display.
Select the type of turbine to be used (reusable or disposable) and press OK. The turbine selection will be saved automatically and the
device will return to the Service Menu.
Turbine calibration
Select the Turbine Calibration item and choose from the following options:
show current values
modify calibration
factory defaults
Selection of the first item shows the percent correction applied in that moment.
The item "modify calibration" allows to insert new calculated values to perform a new calibration. A password is required to access this
option; insert the PIN by touching the numbers; if the PIN is not set digit the default PIN which is the following:
The item "factory defaults" erases the previous calibration values and restores the two percentage corrections to zero percent correction
factor; in this case a password is required as explained above.
To perform this procedure correctly please refer to paragraph 2.7.3.
Select language
Select the desired item by touching the display and press OK, the language is now set and the device will return to the Service Menu.
DATE Format
Select the item by touching the display.
day
month year
month day
year
year
month day
Select the desired format and press the OK icon; the selection will be saved automatically and the device will return to the service menu.
UNIT format
Select the option by touching the display.
Imperial
(in,lb)
Metric
(cm kg
Select the desired format and press OK; the selection will be saved automatically and the device will return to the service menu.