Touch the ESC icon to exit the test, otherwise after a seconds will start the test.
In the event that a test is interrupted due to a complete battery discharge, the next time the device is turned on the following
message is displayed:
At the same time an intermittent beep is emitted for 4 seconds.
Subsequently the Spirodoc returns to the main screen.
WARNING
Avoid twisting the sensor's cable as this may compromise measurement accuracy and the integrity of the sensor itself, also
do not apply excessive force when using, connecting, disconnecting or storing the oximetry sensor.
The first few seconds are used to find the best signal possible; after which the Spirodoc timer resets itself and the device starts recording
data.
For any type of oximetry test if the sensor is not properly connected the following message will be displayed on screen after a few
seconds:
At the same time Spirodoc emits an acoustic alarm (if previously set in the service menu).
If the sensor has been connected properly but the finger has not been properly inserted in the sensor the following message will be
displayed on screen.
At the same time Spirodoc emits an acoustic alarm (if previously set in the service menu).
If the signal reaches the sensor properly, after a few seconds the device will emit the pulse tone an acoustic signal while also displaying
the values on screen.
The alarms can be customized, the procedure is described in paragraph 2.7.1.
During oximetry testing if the SpO2 and/or the Pulse rate fall below the lower threshold or raise above the upper threshold, Spirodoc
will emit an acoustic alarm '(if previously set in the service menu.)' until such situation persists. For sleep oximetry testing the Pulse rate
tone is always disabled.
If all the alarms are activated during oximetry testing the
show up on screen.
By touching this icon during a test the device will display for a few seconds the
alarm settings as can be seen from the image on the right. The alarm thresholds
that have been previously set in the service menu are also visible. After a few
seconds the device returns to the current test screen.
If the
icon shows up during a test, one or more alarms have been disabled in the OFF position in the service menu. The user may
always check the alarm situation by touching the above icon
Upon activating an alarm among those chosen the
an acoustic signal for two minutes. In this case the icon will be shown as
are over.
For information concerning the proper setup of this function please refer to paragraph 3.6.
WARNING
A test is saved with the name of the last patient displayed. If a test refers to a previously saved patient, then prior to performing
a test the user must recall that patient from the database as described in paragraph 3.8.2
WARNING
A test is archived with the name of the last patient displayed; if this relates to a previously entered patient, before carrying out
the test recall a test previously carried out on the subject in question and operate as described at the end of section 2.8.2.
During the SpO2 and 6MWT oximetry tests, the battery pack charge level is shown, allowing an estimate of the available
autonomy, which may vary depending on the state the device is in (display illumination at maximum or in economy mode).
During a test the display will show the following information:
SPIRODOC
WARNING
Wrong interruption of last oximetry test
WARNING
Sensor unplugged
WARNING
Finger not inserted
icon will always
will be visualized. By touching the icon the corresponding alarm will not emit
cod. 980156
Rev 4.2.1
to then return to the previous icon when the two minutes
spirodoc
User manual
EN 39/48