Operator shall not touch simultaneously the patient and the parts of non-medical equipment that are accessible to the
operator during routine maintenance after removal of covers without the use of a tool.
5.
PROBLEM SOLVING
Please find below a list of problems that may arise when using Spirodoc.
Diagnostic messages are also shown on the display indicating the type of malfunction:
PROBLEM
MESSAGE
\
\
Spirodoc does
\
not turn on
Message
Too many
during switch
attempts for
on
today
To comply
with privacy
Message when
policy please
the privacy
update all
info is enabled
connected
devices
Error in ram
Problem when
memory
turning on the
Recovering
device
data
Please wait
\
The device
turns off and
subsequently
turns on again.
\
Spirometry test
results are
\
unreliable
Certain
\
spirometry
and/or
oximetry
parameters are
not shown at
the end of a
test.
During an
\
oximetry test
values are
\
returned at
irregular
intervals,
intermittent or
simply wrong.
During
\
oximetry
testing the
screen is barely
readable
Problem during
Damaged
battery pack
battery pack
recharging
SPIRODOC
POSSIBLE CAUSES
The battery pack could be discharged
The battery pack has not been properly
inserted in the device
The device may have lost its internal
software
The user has exceeded the maximum
number of attempts to insert the pin
The privacy info is enabled.
Memory data within the device has been
damaged
An internal error has occurred.
The turbine may contain dirt or foreign
matter.
The test was not performed correctly.
Personalized parameter setting in the
service menu.
The sensor is positioned incorrectly or the
patient perfusion is insufficient.
The patient has moved.
After a few minutes the screen backlight
turns off automatically to save battery
energy.
The battery pack could be damaged or
simply mispositioned.
cod. 980156
Connect the device to the battery charger.
Contact a technical service center
Connect the device to the PC with the USB cable and
update the internal software; For more detailed
information please consult the MIR Spiro software user
manual available on line within the software itself.
Wait for the next day. The number of attempts will be
resetted.
The message advises the user to update the software
MIR Spiro installed on the PC to comply with privacy
policy
If the data has been restored correctly the standard turn-
on process will complete itself. If this process does not
finalize contact an authorized technical service center.
Check on the following website www.spirometry.com
for a more recent internal software release of the device.
Update the internal software by downloading the latest
release by using the MIR Spiro software For further
information consult the MIR Spiro manual available on
line within the software itself.
Clean the turbine as explained in paragraph 4.1; if
necessary replace the turbine with a new one.
Repeat the test and follow closely the indications shown
on the screen.
Check the parameter setting in the item "PARAMETER
setting" within the Service Menu as explained in
paragraph 2.7
Riposition the oximetry sensor.
To obtain accurate oximetry readings it is important that
the patient must not move abruptly.
None
Contact a technical service center
Rev 4.2.1
spirodoc
User manual
REMEDY
EN 47/48