Bruker S8 TIGER Manual De Instrucciones página 33

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

S8 TIGER X-ray Spectrometer
Warning (applies to both cooling systems)
The pressure of the main water supply must not exceed 6.0 bar (87 PSI). If the local water
pressure is larger, a pressure reducer must be used!
Note (applies to both cooling systems)
The water inlet valves are only opened to the degree necessary. At times they might be
entirely closed. As a result, water consumption is reduced to a minimum.
Assure that all tubing and all fittings outside the instrument are properly dimensioned and
with stand the water pressure.
Check the rating of the components before turning on the instrument.
Note
Reliability and safety of the external plumbing must be assured by the
customer.
DOC-M80-ZXX031 V4 – 02.2010
Introductory User Manual
1-19