S8 TIGER X-ray Spectrometer
Tab. 1-10:
Cooling water supply requirements.
Cooling water supply
(applies to both cooling systems)
Flow rate
Difference pressure
(= difference between input and pressure-free
output pressure)
Water temperature
Maximum particle size (suspended solids)
Tab. 1-11:
Optional water filter.
Water filter (order no. HOT-GWS-2007-02)
Maximum particle size (suspended solids)
Flowrate
Dimensions (wall fastening)
Weight (filled)

Closed-circuit Cooling System (optional external chiller)

If a sufficient city/well water ("tap water") supply is not available, an external cooling water unit with a
closed water circulation loop may be used. In this case, it must be considered that the water flow is not
continuous (see note). The discharge should occur without pressure. Therefore, it may be necessary
to install a bypass.
There are two types of chillers: one dissipates the heat to a secondary water loop, the other to labora-
tory air. Especially if you decide to use the latter version, e.g. the model listed in the table below, we
strongly recommend installing the unit in a room separate from the instrument. This helps to reduce
the overall noise level at the S8 TIGER. The diagram shows the version using a secondary water loop.
DOC-M80-ZXX031 V4 – 02.2010
Technical Data
0 to 15 l/min (0 to 4 gallons/min)
(always adjusted to minimum)
4 to 6 bar (58 to 87 PSI)
10 to 20°C (50 to 68°F) at 6 bar
10 to 16°C (50 to 61°F) at 4 bar
< 90 μm
90 μm
450 l/h
800 x 600 x 170 mm (H x W x D)
16 kg
Introductory User Manual
1-21