S8 TIGER X-ray Spectrometer

System Control

The control panel is located on the front of the S8 TIGER.
Fig. 2-2:
Control panel (including all switches)
DOC-M80-ZXX031 V4 – 02.2010
STOP Button
The STOP button switches off the control electronics, high voltage
generator and all components of the system. Therefore, the X-ray
source will be turned off instantly and all drives will stop their move-
ments.
The STOP button should only be pressed in case of an emergency.
Do not use it for normal shutdown.
Power ON Button
The Power ON button switches on all components of the S8 TIGER,
including the high-voltage generator.
After switching on the system, all four illuminated status LEDs will
flash until the system is fully initialized. The green LED will remain lit,
indicating that the system is ready. After the startup procedure, the
system can be fully activated with the key switch (see below). Sub-
sequently, the instrument enters 'Measurement Standby Mode'.
Power Off Button
The Power OFF button switches off all components of the
S8 TIGER, including the high-voltage generator.
Introductory User Manual
2-11