Introductory User Manual
Note
Configure the output voltage of the Toshiba UPS to 240V!
208V are preset by the manufacturer.
Suitable for the S8-3kW/4kW:
Tab. 4-11:
UPS type Effekta Pegasus 31 (1 ph.)
UPS TYPE Effekta (order no. Pegasus31 -15kVA)
nominal output power
backup time
Dimensions
Weight
Input
Output
Tab. 4-12:
UPS type Toshiba
UPS TYPE Toshiba (order no. UE3G2L140C61T)
nominal output power
backup time
Dimensions
Weight
Input
Output
Note
Configure the output voltage of the Toshiba UPS to 240V!
208V are preset by the manufacturer.
4-8
15 kVA / 12 kW
15 min. at rated power
1200 x 450 x 650 mm (H x W x D)
260 kg
3x32A, 380-415V (±15%), 50/60 Hz
2 x 40 A, 50/60 Hz
14 kVA / 11.9 kW
15 min. at rated power
992 x 445 x 885 mm (H x W x D)
281 kg
2x40A, 208-240V (±10%), 50/60 Hz
2 x 40 A, 240V, 50/60 Hz