Water Chiller For The S8 Tiger (3Kw Or 4Kw Only) - Bruker S8 TIGER Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

S8 TIGER X-ray Spectrometer
Available Toshiba accessories: sub-switch board in-/ output 1-ph, fuses 2x 16 A for 5 safety plugs,
splitter type A02, including cables net-UPS 10 m and UPS-sub-switch board 4 m.
Available Effekta accessories: sub-switch board output Cekon plug socket 3-ph, fuse 1x 40 A, 3 safety
plug power sockets 1-ph, fuse 1x 16 A, splitter type A03, including cables net-UPS 10 m and UPS-
sub-switch board 4 m.

Water Chiller for the S8 TIGER (3kW or 4kW only)

As an option, an external cooling unit with self-contained cooling water circulation loop can be used for
cooling the X-ray generator and X-ray tube. We recommend the following water/air heat exchanger:
Tab. 4-13:
Water Chiller
Water Chiller TYPE 3000 (order no. ERL 3000)
Cooling capacity
Cooling water outlet
Temperature stability
Tank capacity
Water flow
Required electricity
Power consumption
Dimensions
Weight (empty)
Compliance
Degree of protection
The unit is delivered with a 10 m power cable, a 10 m controller cable, and two 10 m water hoses. For
internal pressure compensation, it is equipped with a bypass to improve operation reliability.
specifications can be changed with additional parts.
DOC-M80-ZXX031 V4 – 02.2010
6.6 kW at T
=32°C, T
ambient
15 to 25°C (adjustable)
r1 °C
40 l
300 l/h
400 V, 3-phase, 50 Hz
3.52 kW
1212 x 600 x 741 mm (H x W x D)
220 kg
CE
IP44
Introductory User Manual
= 18°C
cooled water
Technical
4-9