Introductory User Manual
Protection Classification according to IEC 536
Protection Degree according to IEC 529
Electrical Specifications
Tab. 4-2:
Electrical Specifications
X-ray Generator and Control Electronics
Mains Supply Voltage
(Actual nominal voltage is factory-preset and
printed on the system label located on the sys-
tem's rear panel.)
(For details see section 'Pre-Installation - Mains
Connection and Grounding' within this manual)
Frequency Range
Power Consumption
S8-1kW
S8-3kW
S8-4kW
Overvoltage Category according to IEC 664
4-2
x
French radiation safety standard: AFNOR NFC 74-100
Class I
IP 2 0
Three-phase supply:
3 x 120 VAC (+/- 10 %)
3 x 230 VAC (+/- 10 %)
3 x 240 VAC (+6 % / - 10 %)
One-phase supply (on request only):
208 VAC (-10 % / +10 %)
230 VAC (+/- 10 %)
240 VAC (+6 % / -10 %)
47 Hz – 63 Hz
2.8 kVA
5.2 kVA (3-phases mains supply)
7 kVA (3-phases mains supply)
II
S8 TIGER X-ray Spectrometer
DOC-M80-ZXX031 V4 – 02.2010