If there are errors during the test, the test section is marked with
a red X and an appropriate note in the info field is displayed. In
this case, confirm the test result with the sensor button "OK".
The functional test is carried out after any disconnection from
main power.
Note: When pressing the left arrow, a safety interval between
the screen "Basic settings" and "Functional test" is set to pre-
vent accidental starting of the functional test. You may abort the
functional test by immediately pressing the sensor button "C".
3.4.6 Basic settings
Figure 8: Basic settings menu
3.4.6.1 Temperature mode
Pressing the yellow sensor button changes the temperature
mode between °C and °F.
3.4.6.2 Night/Standby temperature
This function is used to set the temperature of the furnace
when switched off but not disconnected from main power.
The night/standby temperature prevents moisture to collect in
the furnace chamber. This temperature is used in the night
mode (see section 4.4) and in standby mode. When switching
the furnace off with the "On/Off" switch, the standby function
is activated automatically.
The yellow sensor button can be used to set the desired
temperature.
1. Press the sensor button (button changes color).
2. Enter the new temperature with the numeric sensor keypad.
3. Temperature >100 °C = night/standby ready
4. Temperature <100 °C = night/standby not ready
5. Press sensor button again (or select next sensor field), and
the new temperature value is saved (sensor field changes
to yellow again).
6. Wrong entries can be deleted with the sensor button "C".
Note:
Several settings of
the Multimat2 Touch
are preset on delivery.
You may accept
these settings by
pressing the sensor
button " " or change
as described here-
after.
3.4.6.3 Vacuum mode
Pressing the yellow sensor button changes the vacuum mode
between hPa, inHg and Hg".
In the modes inHg and Hg", the pressure is displayed with one
decimal.
Note:
R
hPa
1013 hPa = standard pressure at sea level
R
inHg
29.9 inHg = standard pressure at sea level
R
Hg"
0 Hg"
= standard pressure at sea level
(US standard)
3.4.6.4 System time
1. Press the sensor button next to "Time"
(button changes color).
2. Enter the 4-digit number for hours/minutes with the numeric
sensor keypad.
3.4.6.5 Date
1. Press the sensor button next to "Date".
2. Enter the 8-digit number for day/month/year with the
numeric sensor keypad.
3.4.6.6 Acoustic signal
Press the sensor keypad to switch the acoustic signal on or off.
3.4.6.7 Idle temperature
This function is used to save energy and reduce the surface
temperature of the furnace.
Note: The idle temperature is factory-set to 400 °C. It cannot be
set to temperatures > 600 °C. The idle temperature must be set
to 25 °C lower than the preheating temperature.
1. Press the sensor button next to "Idle temperature".
2. Enter the new temperature with the numeric keys of the right
control pad.
3.4.6.8 Upper temperature limit
This function is used to limit the maximum permissible tempera-
ture of the furnace.
Note: It is recommended to limit the upper temperature value,
if a heavy stress to the muffle can and shall be prevented.
1. Press the sensor button next to "Upper temperature limit".
2. Enter the new temperature with the numeric keys of the right
control pad.
3. Press the sensor button "Upper limit value" again (or select
next sensor field), and the new temperature value is saved
(sensor field changes to yellow again).
3.4.6.9 Data output
By pressing the sensor button, the data output changes between
"printer", "PC", and no output.
If the symbols "printer" or "PC" are activated, all target and
actual values are automatically transferred to printer/PC at the
end of each firing cycle, as long as a printer or PC is connected.
47