6.16.1 Program transfer furnace
1. Insert the Smart Media Card into the slot on the right rear
side of the furnace (metal surface facing back).
2. Press the sensor button "M" on the numeric keypad.
3. Select "Smart Card".
4. Press the sensor button "Furnace
5. Select individual programs from the list or press "All" in the
lower soft key bar.
R
6. Press "
Card" in the lower soft key bar.
7. Confirm program saving by pressing "C".
6.16.2 Program transfer Smart Media Card
Using this function, you can transfer programs from the
Smart Media Card to the Multimat2 Touch.
1. Insert the Smart Media Card into the slot on the right rear
side of the furnace (metal surface facing back).
2. Press the sensor button "M" on the numeric keypad.
3. Select "Smart Card".
4. Press the sensor button "Smart Card
5. Confirm loading programs with "S" = yes
(abort with "C" = no).
6. The program(s) is/are being loaded.
7. Confirm program saving with the "C" button.
Service note:
In the event that your firing programs for older Multimat models
were previously stored on a Smart Card (see Figure 28) and you
want to continue using your data on a Multimat2, you can have
the data transferred to a Smart Media Card, free of charge, by
your distributor.
Figure 28: Smart Card, front and rear view
R
Smart Media Card
R
Smart Card".
R
furnace
R
Furnace".
7 Service and maintenance
7.1 Lift
The lift is used for vertical transport of the firing chamber. The
unit is driven by a geared motor. If there is a power failure, the
firing chamber can be lifted upwards by hand in order to remove
the work pieces. Apply vigorous pressure downwards to move
the firing chamber downwards by hand.
Caution! Do not change the setting of the belt tension screw.
A too strong belt setting may cause overload of the lift drive,
thus damaging the unit.
7.2 Firing socket/pressing socket and Firing socket
carrier
The firing platform carrier is mounted vertically resilient on the
top of the furnace subsection. The O-ring in the firing socket
carrier seals the firing chamber during vacuum firing. Depending
on whether the unit fires or presses, the firing socket or the pres-
sing socket is placed on the firing socket carrier. Both firing and
pressing socket insulate the firing chamber against downward
temperature loss and are also used as object carrier.
Maintenance note: The O-ring of the firing socket carrier must
be kept clean and should be occasionally checked for damage.
Caution! In order to prevent damage to the firing socket carrier
and the O-ring seal, either the firing socket or the pressing socket
must be placed on the firing socket carrier.
7.3 Vacuum pump
We recommend operating the Multimat 2 Touch with the
highperformance Dentsply vacuum pump.
However, the Multimat2 Touch can be operated with any
highperformance vacuum pump with grounded plug connector
(power uptake max. 2.5 A) that has a liter performance of at
least 30 l/min and a final pressure of 30 hPa. Connecting the
pump is described in section 3 "Setup and first operation".
The pump connection tube must not be longer than 2 meters
Maintenance note: Observe the operating manual of the pump
for maintenance purposes.
Important: Vacuum pumps with oil lubrication must have an oil
change every 3 months.
61