Uline OHAUS RANGER 7000 Manual Del Usuario página 35

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

ERROR CODE
EEP Error
Power on Overload
Overload
Underload
Tare Error
Display Overflow
No Calibration
-----------
---NO---
Calibration Error
Low Reference
Reference Error
If the troubleshooting section does not resolve your problem, contact ULINE Customer Service at 1-800-295-5510.
PAGE 35 OF 105
TROUBLESHOOTING
DESCRIPTION
EEPROM checksum error
Power on error
Over range error
Under range error
Tare out of range error
Display overflow
Calibration data error
Busy message
Action not allowed message
Calibration error
Low reference weight
warning message
Unacceptable reference
weight message
1-800-295-5510
Corrupted EEPROM.
Weight reading exceeds Power On Zero Limit.
Weight reading below Power On Zero Limit.
Weight reading exceeds Overload Limit.
Weight reading below Overload Limit.
Tared at one unit, but after switching to another
unit, the tare value exceeds the maximum.
Weight exceeds six digits.
Calibration data does not exist.
Displayed during tare setting, zero setting, printing.
Function not executed.
Calibration value outside allowable limits.
Average Piece Weight too small (Warning).
Reference Weight too small. The weight on the pan
is too small to define a valid reference weight.
uline.com
CAUSES
0721 IH-7991

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

H-7991H-7992H-7993H-7994H-7995

Tabla de contenido