Aesculap
ELAN 4 electro Standard handpiece (1 ring) GA861 to GA864 / ELAN 4 electro
HD handpiece (2 ring) GA865 to GA869
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
D–W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
►
Connect the connector for motor cable 3 with ELAN 4 electro rinsing
device GB692R.
►
Place the rinsing device on a wire basket suitable for cleaning.
►
Connect the rinsing device to the rinsing connection of the rinsing cart.
►
After the mechanical cleaning/disinfection process:
– Remove any residual rinsing water from the applied part using a
compressed-air gun, see instructions for use ELAN 4 rinsing device
and rinsing adapter TA014447 and TA014448.
– Check visible surfaces for residues.
12
®
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD–W
>10/50
1
FD–W
90/194
5
FD–W
-
-
-
Chemicals
-
■
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
■
0.5 % working solution
– pH ~ 11*
-
-
In accordance with the program for the cleaning and disinfecting machine