4.
Product description
4.1
Available sizes
Note
The shaft length can be seen on the shaft of the drill handpiece.
ELAN 4 electro Standard handpiece (1 ring)
Art. no.
Length of shaft
GA861
L4 (approx. 4 cm)
GA862
L7 (approx. 7 cm)
GA863
L10 (approx. 10 cm)
GA864
L13 (approx. 13 cm)
ELAN 4 electro HD handpiece (2 ring)
Art. no.
Length of shaft
GA865
L7 (approx. 7 cm)
GA866
L10 (approx. 10 cm)
GA867
L13 (approx. 13 cm)
GA868
L17 (approx. 17 cm)
GA869
L22 (approx. 22 cm)
4.2
Package includes
ELAN 4 electro Standard handpiece (1 ring)
Art. no.
Designation
GA861–GA864
ELAN 4 electro Standard handpiece (1 ring)
TA014443
Instructions for use of ELAN 4 electro Standard
handpiece (1 ring) GA861 to GA864 /
ELAN 4 electro HD handpiece (2 ring) GA865 to
GA869 (leaflet)
ELAN 4 electro HD handpiece (2 ring)
Art. no.
Designation
GA865–GA869
ELAN 4 electro handpiece HD (2-Ring)
TA014443
Instructions for use of ELAN 4 electro Standard
handpiece (1 ring) GA861 to GA864 /
ELAN 4 electro HD handpiece (2 ring) GA865 to
GA869 (leaflet)
4.3
Components required for operation
Operation-ready ELAN 4 electro control unit GA800, see TA014401
ELAN 4 tool (1 ring or 2 ring)
4.4
Operating principle
Note
The ring marking 2 shows the shaft diameter of the tool to be used. ELAN 4
tools can be only coupled to the applied part if the ring markings
(1 ring/2 ring) of the applied part and tool match.
The drill handpiece is fitted with an ELAN 4 tool coupling (1 ring or 2 ring).
It is used with ELAN 4 electro foot controls/manual controls.
With the drill handpiece, the coupled tool rotates at the set motor speed.
The motor speed of the drill handpiece can be adjusted continuously by
the manual/foot control.
The drill handpiece can be operated in right or left (clockwise or counter-
clockwise) rotation mode.
5.
Preparation
Non-compliance with these rules will result in complete exclusion of lia-
bility on the part of Aesculap.
Do not use products from open or damaged sterile packaging.
Prior to use, inspect the product and its accessories for any visible
damage.
Use the products and their accessories only if they are in perfect tech-
nical condition.
5