Attachments for small drill GA344 and drill GA844
7.2
General information
Dried or affixed surgical residues can make cleaning more difficult or inef-
fective and lead to corrosion. Therefore the time interval between appli-
cation and processing should not exceed 6 h; also, neither fixating pre-
cleaning temperatures >45 °C nor fixating disinfecting agents (active
ingredient: aldehydes/alcohols) should be used.
Excessive neutralizing agents or basic cleaners may result in a chemical
attack and/or fading and the laser marking becoming unreadable either
visually or by machine.
On stainless steel, residues containing chlorine or chloride (such as surgi-
cal residues, drugs, saline solutions in water for cleaning, disinfection and
sterilization) may lead to corrosion (pitting corrosion, tensile corrosion)
and thus to the destruction of the product. These must be removed by
rinsing thoroughly with demineralized water and then drying.
Perform additional drying, if necessary.
Only process chemicals that have been tested and approved (e.g. VAH or
FDA approval or CE mark) and which are compatible with the product's
materials according to the chemical manufacturers' recommendations
may be used for processing the product. All the chemical manufacturer's
application specifications must be strictly observed. Failure to do so can
result in the following problems:
Visual material changes (such as fading or color changes in titanium or
aluminum). For aluminum, the application/process solution only needs
to be pH >8 to cause visible surface changes.
Material damage (such as corrosion, cracks, breaks, premature aging or
swelling).
Do not use metal cleaning brushes or other abrasives that would dam-
age the product surface and could cause corrosion.
For further detailed information on hygienically safe and material-pre-
serving/value-preserving reprocessing, see www.a-k-i.org, link to Pub-
lications, Red Brochure – Proper maintenance of instruments.
7.3
Preparations at the place of use
Remove all attached components from the product (tool and accesso-
ries).
Detach the attachment from the drill.
Remove any visible surgical residues as much as possible with a damp,
lint-free cloth.
Place the dry product in a sealed waste container and forward it on for
cleaning and disinfection within 6 hours.
7.4
Preparation before cleaning
Before the first automatic cleaning/disinfection: attach ECCOS holders
in a suitable tray.
Insert the products in the correct position into the ECCOS fixations, see
Fig. E.
Kirschner wire attachment GB894R
Set adjustment sleeve to the largest Kirschner wire diameter.
10
Acculan 4 / ELAN 4 electro
®
7.5
Cleaning/disinfection
7.5.1
Product-specific safety instructions for the reprocessing pro-
cedure
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
CAUTION
Use cleaning and disinfecting agents according
to the manufacturer's instructions which
– are approved for plastic material and high-
grade steel,
– do not attack softeners (e.g. in silicone).
Do not use cleaning agents that contain ace-
tone.
Observe specifications regarding concentration,
temperature and exposure time.
Do not exceed the maximum temperature of
60 °C during chemical cleaning and/or disinfec-
tion.
Do not exceed the maximum temperature of
96 °C during thermal disinfection with FD
water.
Dry the product for at least 10 minutes at a
maximum of 120 °C.
Note
The indicated drying time is a guide time only. It must be checked taking
into account the specific conditions (e.g. load) and if applicable adjusted.