Aesculap
Attachments for small drill GA344 and drill GA844
6.
Working with the device
Risk of infection and contamination!
Product is delivered unsterilized!
WARNING
►
Sterilize the product before use pursuant to the
operating instructions.
Risk of injury and material damage due to acciden-
tal activation of the product!
►
Products which are not being actively used must
WARNING
be secured against accidental activation (posi-
tion OFF).
Risk of injury and material damage due to inappro-
priate use of tools!
►
WARNING
Always follow the safety advice and information
given in the instructions for use.
►
When coupling/uncoupling, handle tools with
cutting edges with care.
Damage to the product if dropped!
►
Use the products only if they are in perfect
technical condition, see Function check.
WARNING
Risk of burns to skin and tissue through blunt tools
or if product has not been maintained properly!
►
Use tools only if they are in perfect condition.
WARNING
►
Replace blunt tools.
►
Maintain the product properly, see maintenance
guide.
6.1
System set-up
6.1.1
Connecting the accessories
Risk of injury due to unapproved configuration
using additional components!
►
Ensure that the classification of all components
DANGER
used is consistent with the classification of the
product (such as type BF or type CF).
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
6
Acculan 4 / ELAN 4 electro
®
All
configurations
must
IEC/EN/DIN 60601-1. The person connecting the devices with each other
is responsible for the configuration and must ensure that the fundamental
standard IEC/EN/DIN 60601-1 or relevant national standards.
►
Follow the instructions for use of individual accessories.
►
Please address your B. Braun/Aesculap partner or Aesculap Technical
Service with any inquiries in this respect; for a contact address, see
Technical Service.
6.1.2
Protection against inadvertent activation
To prevent the drill being inadvertently activated when changing a tool or
accessory, the knob for speed control can be locked.
Locking the knob for speed control 2:
►
Twist the safety catch 3 to position OFF.
The speed control knob 2 is blocked and the product 1 cannot be oper-
ated.
Unlocking the knob for speed control 2:
►
Twist the safety catch 3 to position ON.
The speed control knob 2 is unlocked and the product 1 can be oper-
ated.
Note
For more information on the small drill GA344, see TA014550 and
TA014551 (flyer).
For more information on the drill GA844, see TA014436 and TA014437
(flyer).
6.1.3
Attaching/removing attachment for drill
Risk of injury when attaching/removing attach-
ments/tools in the ON position through inadvertent
activation of the product!
WARNING
►
Only attach/remove attachments/tools in the
OFF position.
►
Secure drill 1 from inadvertent activation with the safety catch 3, see
Protection against inadvertent activation.
Coupling
Note
Aesculap recommends coupling the Reciprocating Saw attachment 24
with the label "TOP" upward, aligned to the drill.
►
Push attachment 6/9/12/16/18/24 onto the drill 1 until it clicks into
place.
►
Pull on attachment 6/9/12/16/18/24 to check secure connection.
fulfill
the
fundamental