7.7.2
Mechanical alkaline cleaning and thermal disinfection
Machine type: single-chamber cleaning/disinfection device without ultrasound
Phase
Step
I
Pre-rinse
II
Cleaning
III
Intermediate rinse
IV
Thermal disinfection
V
Drying
D-W:
Drinking water
FD–W:
Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least)
*Recommended: BBraun Helimatic Cleaner alkaline
►
For Kirschner wire attachment GB894R: Set the adjustment sleeve to
the greatest Kirschner wire diameter.
►
Insert the product in the correct position into the ECCOS fixation, see
Fig. E.
►
Then connect the interior rinse device to the ECCOS holder and connect
to the rinse connector of the cleaning/disinfection device/rinse cart.
►
After automatic cleaning/disinfection, check visible surfaces for resi-
dues and repeat the cleaning/disinfection process as needed.
T
t
Water
[°C/°F]
[min]
quality
<25/77
3
D–W
55/131
10
FD–W
>10/50
1
FD–W
90/194
5
FD–W
-
-
-
Chemistry/Note
-
■
Concentrate, alkaline:
– pH ~ 13
– <5 % anionic surfactant
■
0,5 % working solution
– pH ~ 11*
-
-
mind. 10 min at max. 120 °C